Working example:



next up previous contents
Next: Flat-field accuracy (LW) Up: Transients and memory Previous: Transients and memory

Working example:

for the LW channel (Fig gif) : the procedure is exactly the same as for SW, only, since = 0, the curves merge together when TO FROM, and they are horizontal when . If FROM = 30 ADUs and TO = 10 ADUs : = 36. If FROM = 30 ADUs and TO = 100 ADUs : = 28.

The experimental numbers are within a factor of 2 of this fit.

On Figure gif, it can be seen that after readouts the image in some cases is still somewhat distorted.

Stabilization after changes of flux or integration time is the main problem encountered with this array. The worst situations occur in two cases, a) illuminating the array previously at dark, or b) after a saturating flux.

The first problem can be overcome by keeping always light on the array and sorting the sequences of observations by decreasing fluxes. Dark calibration must be done rarely, and must be preferentially placed at the end of the observing sequence.

After saturation, the array keeps memory of the illuminating scene for a period as long as half an hour. The only way we found to avoid polluting the following observations with the memory image is to saturate the whole array with a uniform illumination. With an ICD flash of a couple of readouts, the array looses memory of the previous scene, and is ready to image a new field. The drawback of this method is that after the flash, the overall responsivity of the array will be below the average value for about 10 minutes. This effect will decrease the absolute photometric accuracy, but will keep a reasonable relative photometric accuracy inside an image.



ISOCAM Observer's Manual - V1.0
Tue Oct 31 12:06:23 MET 1995