Maintainer: babar@ipac.caltech.edu
A. Use the ISO
archive.
A. You will need the CAM software to get started.
See Data Reduction
and Analysis software for CAM
A. There are several documents available from the IDC
at VILSPA
General CAM information:
http://www.iso.esac.esa.int/manuals/HANDBOOK/III/cam_hb/Dark current:
The ISOCAM Dectectors: An Overview
http://www.iso.esac.esa.int/users/expl_lib/CAM/darkdoc.ps.gzFlat Fields:
Report on the Behaviour of ISOCAM LW Darks (Roman & Ott 1999)
http://www.iso.esac.esa.int/users/expl_lib/CAM/flatdoc.ps.gzGlitches:
Glitch Effects in ISOCAM Detectors (Claret et al. 1998)Transients:
The ISOCAM glitch libraray (Claret & Dzitko 1998)
Cosmic Ray effects on the ISOCAM Long Wave detector (Dzitko et al. 1998)
The Transient Behaviour of the Long Wavelength Channel of ISOCAM (Abergel et al. 1998)
Transient correction of the LW-ISOCAM data for low contrasted illumination (Coulais et al. 1999)
About Transients, Technical Report (Delattre 1996)
Modelling the transient effects in the IR array of the SW channel of ISOCAM (Tiphene et al. 1999)
Flux Calibration:
CAM calibration explanatory document (Altieri et al. 1998)
ISOCAM Photometry Report, V 2.1.1 (Blommaert 1998)
ISOCAM Faint Source report (Altieri et al. 1998)
CAM field of view distortions:
ISOCAM LW Channel Field of View Distortion (Draft) (Aussel 1998)CAM PSF
ISOCAM PSF Report (Okumura 1998)
CAM stary light ("ghosts")
Ghosts in ISOCAM images (Okumura et al. 1998)
A. The List of Contacts
General information on ISO at IPAC: iso@ipac.caltech.edu
CAM: babar@ipac.caltech.edu
[ BACK ]